Home Precision Environment Aerosol Conditioning DDS 560/C Dynamic Dilution System
DDS 560/C Aerosol Dilution System product image thumbnail
DDS 560/C Aerosol Dilution System workflow diagram thumbnail
Topas GmbH Aerosol Dilution Systems

DDS 560/C Dynamic Dilution System

Touchscreen-controlled dynamic aerosol dilution system with automatic readjustment, variable 0.5–3.0 L/min sample flow and RS-232 integration for automated aerosol measurements.

Dilution range
5–370, flow-dependent
Sample flow
0.5–3.0 L/min
Interface
Touchscreen and RS-232
Control
Automatic readjustment
Touchscreen RS-232 Automatic control VDI 3491-15
VDI 3491-15 VDI 3491-6

Model Overview

Model Configuration Primary use Selection note
DDS 560/C 5–370, flow-dependent; 0.5–3.0 L/min Touchscreen-controlled dynamic aerosol dilution system with automatic readjustment, variable 0 Select DDS 560/C for automated, remotely monitored dilution with automatic setpoint control and variable instrument sample flow.

Key Features

Wide adjustable dilution range

Provides a freely adjustable dilution factor over approximately one decade, with the available range dependent on sample flow.

Automatic setpoint readjustment

The control system readjusts the dilution setting to maintain the selected value as operating conditions change.

Variable passive sample flow

Compatible with aerosol analytical instruments using sample flow rates between 0.5 and 3.0 L/min.

Touchscreen operation

Local touchscreen menus support setup, operation and instrument-status review.

Remote communication

RS-232 communication supports remote control and parameter monitoring in automated measurement systems.

Fast operating setup

Published setup time is no more than 10 seconds under the specified operating conditions.

Technical Specifications

Parameter Unit Value / Description
DILUTION PERFORMANCE
Setting parameter Dilution factor
Setting range 5–200 at 0.5 L/min; 10–370 at 1.0 L/min; 30–190 at 3.0 L/min
Setting resolution Steps of 1 or 10
Inlet volumetric flow rate L/min 0.5–3.0, continuously variable
Setup time s Maximum 10
CONTROL & CONNECTIONS
Power supply 24 V DC via power adapter
Operating medium Air; other media on request
Communication RS-232, M9 IP67 5-pole
User interface Touchscreen
Power consumption W Maximum 25
Hose connector mm Ø 8 outer diameter
PHYSICAL
Dimensions (W × H × D) mm 140 × 285 × 200
Weight kg 3.3

Specifications follow the current Topas GmbH product information reviewed for this page. Confirm the final configuration, accessories and calibration documentation at quotation stage.

Applications

Aerosol-instrument calibration

Provides defined concentration reduction for calibration and verification of aerosol analytical instruments.

Indoor aerosol analysis

Conditions indoor aerosol samples before particle counting or size-distribution measurement.

Workplace aerosol measurements

Extends the useful concentration range of instruments used for occupational aerosol monitoring.

Fundamental aerosol research

Allows automated and repeatable variation of aerosol concentration in experimental studies.

Remote-controlled test systems

Integrates with automated benches through RS-232 communication and monitored operating parameters.

Coincidence-error reduction

Reduces concentrated aerosol samples before optical particle counting or mobility analysis.

Similar Products